Still deciding? Get samples of $ !
Order Sample
|
| Customization: | Available |
|---|---|
| After-sales Service: | Online Video Service |
| Warranty: | 1years |
| Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
|---|
| Payment Methods: |
|
|---|---|
| Support payments in USD |
| Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
|---|
| Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
|---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency


Organiser | |
Plunger in-line cleaning | standard, high/medium/low speeds |
Binary high pressure infusion pump | |
Maximum pressure resistance | 9000 psi (62Mpa) |
Flow rate range | 0.001 -10.000 mL/min (increments of 0.001mL/min) |
Flow rate accuracy | ≤0.06% RSD (ASTM) |
Flow rate accuracy | ±0.2% |
Pressure pulsation | ≤ 1% |
Accuracy of gradient | ±0.5% |
Gradient repeatability | ≤0.1% |
Plunger online cleaning | standard, high / medium / low speed |
UV-visible detector | |
Wavelength range | 190-800 nm |
Light source | deu terium lamp, tungsten lamp |
Spectral bandwidth | 8 nm |
Wavelength accuracy | ±1 nm |
Noise | ±0.35×10-5 AU (ASTM) |
Drift | 1×10-4 AU/Hr |
Wavelength accuracy | 0.1 nm |
Linearity range | >2.5 AU (ASTM) |
Maximum sampling rate | 100Hz |
Flow cell withstanding pressure | 1200PSI |
Flow cell optical range | 10mm |
Minimum detection concentration | 4×10-9 g/mL |
Autosampler | |
Sampling mode | full ring injection/partial injection/microlitre carryover injection |
Volume of quantitative ring | 100 μL as standard |
Sampling repeatability | full loop injection: RSD ≤ 0.3%; partial injection: RSD ≤ 0.5%; microlitre carry-in injection: RSD ≤ 1%. |
Sample volume | 2 standard 48-bit sample bottle holders 1.5mL sample bottle (standard) |
(Optional 96-well plate, 384 small well plate) | |
Cross-contamination | <0.005% |
Linearity | 0.9999 |
Maximum pressure | 6000psi (can be customised) |
Column Temperature Chamber | |
Temperature control range | room temperature +5 ℃-70 ℃ |
Temperature Accuracy | ±1.0°C |
Temperature stability | ±0.1℃ |
Column capacity | 250mm chromatographic column 1 (can be added with a 10mm protection column) |